STATIM Water Reservoir Filter

Update

Please be aware of the Statim water reservoir filter that will be included to all new Statim 2000(S)(E)(G4) and 5000(S)(G4) effective April 2006 as well as with all replacement filler cap spare part kits.

GET the Spare Parts Kit: Reservoir Cap / Water Reservoir Filter

This filter is available as a stand-alone spare part for regular maintenance.

Please note the filter will not work with old reservoir caps and therefore ordering a new reservoir cap is recommended to upgrade existing field units. All replacement caps will come with a new water reservoir filter.

  • The water reservoir filter will greatly reduce the occurrence of particles entering the water reservoir and help prolong the life of the internal components
  • Backward compatible with older units when a new cap is installed
  • Avoids splashing when pouring distilled water into the reservoir
  • The filter is easily removable for monthly cleaning or when required, due to a clogged filter, by placing the filter upside down below running water

SciCan strongly recommends the water reservoir filter become a part of all Statim maintenance and repairs.

How to replace/clean the Water Reservoir Filter

Pump 110V Statim 2000

01-100568S
$278
In stock
1
Product Details
Brand: SciCan

Water pump. For STATIM 1000 & 2000 units, 120V.

Water Pumps

STATIM units have been manufactured using three different pumps. Before servicing, you must first determine which model of pump the unit contains (See Figure 1). Important Note: Always make sure that there is sufficient steam-process distilled water in the reservoir prior to testing STATIM pumps.

Identifying Pump Types

Fluid-O-Tech Pump

The Fluid-O-Tech pump is generally factory fitted to older units with revision 3.x/4/x controller boards up to approximately 1994-1995. It is a small rectangular pump with a diode plug attached to the top. The pump rests on a vibration-damping pad on the bottom of the chassis and is held in place with a wire bracket. The pump is secured to the bracket by a cable tie at each end. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

SciCan Pump

The SciCan pump is generally factory fitted to units with revision 2.x/5.x/6.x and some early 7.x controller boards from 1995 up to mid 2008. It has a similar footprint to the Fluid-O-Tech pump, however the diode is now integrated into the pump. The pump is mounted on two vibration damping rubber brackets. Each bracket is secured to the chassis with two screws. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

Ulka Pump

The Ulka pump is factory fitted to units with revision 7.x controller boards only from mid 2008 onwards. It is slightly smaller than the SciCan pump and can be identified by the configuration of the wires on top of the pump. The two visible connections are located adjacent and parallel to each other whereas the two visible connections on the SciCan pump are opposing each other and have a thermal switch connecting them. The pump is mounted on two vibration damping rubber brackets. Each bracket is secured to the chassis with two screws. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

Note: Important information on pump interchangeability:

  • The Fluid-O-Tech pump is no longer available, but units with this pump can be retrofitted with the SciCan pump ONLY.
  • The SciCan pump and Ulka pump are NOT interchangeable. They are both still available. If replacing the pump during maintenance or servicing, the same pump must be fitted to ensure the unit works correctly.