STATIM seals go green!

Replacing the Cassette Seal with a new green Statim seal

 

Update

New production of Statim 5000 units and replacement seals will be changing this week followed by the Statim 2000 units and replacement seals within the next 3-4 months. The Statim 7000 units and seals should be changing by the end of the year. The dates may change depending on inventory levels at the SciCan warehouses.

Now SciCan quality is easier to recognize.

Pump Recovery kit 110V Statim 2000

SKU 01-108800S
$27
Pre-order
LENGTH
Please choose
Out of stock, available for pre-order
1
Save this product for later
Have questions?
Share this product with your friends
Customer reviews
Reviews only from verified customers
No reviews yet. You can buy this product and be the first to leave a review.
Pump Recovery kit 110V Statim 2000
Product Details
Brand: SciCan
Weight: 0.02 lbs

STATIM PUMP RECOVERY KIT

for Water pumps. For STATIM 1000 & 2000 units, 120V.

Performing Pump Tube Replacement

The pump tube located between the pump and the steam generator is designed to control the flow of water into the steam generator so that it receives the precise amount of water required at specific points in the sterilization cycle.
The tube is directional (pump to steam generator) and contains a precision orifice tube inside the main tube that has a precise diameter that controls the volume passing through into the steam generator. The orifice tube is located at the steam generator end of the main pump tube. The diameter of the fitting is represented on the outside of each pump tube by a numeric value. The larger the value, the larger the diameter of the orifice.
If the filters are clean and the pump flow parameters are not correct, then it may be possible to recover the correct water delivery by the replacement of the pump tube with one of larger or smaller diameter as appropriate.
Single pump tubes are available in a large number of diameters from 0.0210” diameter to 0.0290” diameter in 0.0005” increments.

Note: The number as it appears on the tube may only be represented by 3 digits, e.g. a 0.0245” tube will be represented by ‘245’.