Product Support
Description:
Internal cleaning of Backflow Preventer. This procedure applies to All HYDRIM® G4 WD units C61WD & M2 G4 for all markets except North America.
The HYDRIM C61w G4 is equipped with an air gap / anti-suction device to prevent
backflow of dirty water into the water supply. No other air gap device is necessary.NOTE: When a cycle is started (P1, P2, P3) the air gap is filled four times before the circulation pump starts.
Referenced Parts
Backflow Preventer 01-113318S (C61WD) and 01-113853S (M2 G4).
Required Materials
The following additional materials are required for this step:
- 1x large syringe with tube (or sprayer bottle)
- 1x citric acid-based instrument neutralizer (for example, Dr. Weigert Neutralizer or distilled white vinegar)
- A long flexible bottle brush may be helpful
Cleaning Procedure: Backflow Preventer/Air Gap
- Remove the top cover from the HYDRIM unit and the left side panel (C61WD) or rear panel (M2 G4).
- Fill a large syringe with 50ml neutralizer and then inject it through the opening on the Air Gap (Fig. 1).
- Turn on the device and go into the technician menu: Technician menu / Diagnostic tools / Component test / Filling Draining (Fig. 2).
- IMPORTANT: Start the filling process by pressing the “On” button. Once the unit reaches the position “E = On, F=On,” immediately press the “On” button again. This action will prevent the unit from activating the Air Gap pump and emptying the air gap.
- WARNING: The device may still contain residues of dirty water. Be sure to drain them via the drain hose as described, and do not pump into the wash chamber by starting the Air Gap pump. Otherwise, these contaminants can enter the pump and damage it.
- Leave the liquid to work for approximately 10 minutes. Tap gently on the side of the Backflow Preventer with the palm of your hand so that the agent spreads and the dirt dissolves.
- Drain the liquid via the drain hose (at the bottom of the unit). Repeat the process with the filling from point 2 on, for further visible contamination (e.g., shadows, coverings, biofilm).
- When no more contamination is visible, apply another 25ml neutralizer with the syringe and inject it into the air gap via the opening. Fill the Backflow Preventer again by going into the Technician Menu / Diagnostics Tools / Component Testing / Filling Draining. This time, do not press the “On” button again after reaching the F-switch position. This allows the Air Gap pump to start, which cleans and descales it as it empties the air gap of the solution. Repeat this step if necessary.

Fault 02
- Timeout on filling/emptying up the Air Gap reservoir/chamber. C61 – 150s. For RO – 120s. L110 – 200s. For RO – 600s. M2 – 180s. For RO – 200s
Probable Cause
- Water supply issue
- Water valves failure
- Air gap water pump failure
- Air gap valve failure
- Air gap Full/Empty level switches failure
- Chamber water level pressure switch malfunction
- Overflow switch malfunction triggering evacuation pump
Tech Troubleshoot
- Water supply issue :water valves, water pump or air gap valve have failed.
- Air gap Full/Empty level switches failed.
- Chamber water level pressure switch malfunction.
- Overflow switch malfunction triggering evacuation pump.
User Troubleshoot
- Contact dealer service department.
Sources:
© 2019 SciCan Ltd., SciCan is a COLTENE Group Company. All rights reserved.
www.scican.com SciCan Ltd., 1440 Don Mills Road, Toronto, ON M3B 3P9, Canada
Washers
Don’t Get Stuck in the Steri-Center
HYDRIM® get you back chairside, quickly and safely.
HYDRIM Compact: 6 full-size cassette capacity allows for 120 instruments turnover, in as little as 44 minutes, start to dry. HYDRIM Large: 10 full size cassette capacity allows for 200 instrument turnover, in as little as 44 minutes, start to dry.
StatClean ultrasonic efficacity
Refine by
Excellent for New Jewelry

Ultrasonic Cleaner 800ml/0.85quart
List price $139.99
Save $9.99
$130
Ultrasonic Cleaner 800ml/0.85quart
Convenient beaker holder, holds single use PET cups, or a variety of instruments or jewelry to complement your cleaning needs.
Powerful cleaning

Ultrasonic Cleaner 6L/1.3gal
$720
Ultrasonic Cleaner 6L/1.3gal
Replaces the StatClean! Tight lid to prevent aerosol escape. No need for an additional protective cover.
Powerful cleaning

Top of the Line Ultrasonic Cleaner for Piercing, Tattoo, dental, and surgical instruments, 1.6Gal/6L
$899.95
Top of the Line Ultrasonic Cleaner for Piercing, Tattoo, dental, and surgical instruments, 1.6Gal/6L
Top of the line! Tight lid to prevent aerosol escape. No need for an additional protective cover.
Powerful cleaning

Top of the Line Ultrasonic Cleaner for Piercing, Tattoo, veterinary, and surgical instruments, 2.1 Gal/8L
$1,199.95
Top of the Line Ultrasonic Cleaner for Piercing, Tattoo, veterinary, and surgical instruments, 2.1 Gal/8L
Top of the line! Tight lid to prevent aerosol escape. No need for an additional protective cover.
Excellent for Jewelry

Ultrasonic Jewelry and Eyewear Cleaning Solution Concentrate
$10
Ultrasonic Jewelry and Eyewear Cleaning Solution Concentrate
Jewelry cleaning concentrate to prepare for anodization and sterilization