STATIM Water Reservoir Filter

Update

Please be aware of the Statim water reservoir filter that will be included to all new Statim 2000(S)(E)(G4) and 5000(S)(G4) effective April 2006 as well as with all replacement filler cap spare part kits.

GET the Spare Parts Kit: Reservoir Cap / Water Reservoir Filter

This filter is available as a stand-alone spare part for regular maintenance.

Please note the filter will not work with old reservoir caps and therefore ordering a new reservoir cap is recommended to upgrade existing field units. All replacement caps will come with a new water reservoir filter.

  • The water reservoir filter will greatly reduce the occurrence of particles entering the water reservoir and help prolong the life of the internal components
  • Backward compatible with older units when a new cap is installed
  • Avoids splashing when pouring distilled water into the reservoir
  • The filter is easily removable for monthly cleaning or when required, due to a clogged filter, by placing the filter upside down below running water

SciCan strongly recommends the water reservoir filter become a part of all Statim maintenance and repairs.

How to replace/clean the Water Reservoir Filter

Leveler

01-104180S
$8.50
In stock
1
Product Details
Brand: SciCan
Model: 2000, 5000, 7000, 900

Replacement leveler foot for Statim autoclaves


Replacement leveler repair kit and foot

INSTRUCTION FOR LEVELER REPAIR
1. REMOVE BROKEN LEVELER INSERT
2 IF NECESSARY DEBURR EDGES FROM THE CHASSIS. THE LEVELER MOUNTING HOLE
3. MARK POSITION OF TWO 3/32 [2.5mm] HOLES FOR T-NUT REPAIR INSERT MOUNTING SCREWS (HOLES ARE 600 15.25mm APART). FOR MARKING USE T-NUT AS A TEMPLATE.

4 INSTALL T-NUT REPAIR INSERT FROM THE BOTTOM OF THE CHASSIS AND SECURE WITH TWO INCLUDED SCREWS WITH LOCKWASHERS AND NUTS INCLUDED M2.3 USE OF LM113 LOCTITE (OR EQUIVALENT) IS HIGHLY RECOMMENDED.

5 IN CASE OF REPLACING LEVELER INSERT HIDDEN UNDER THE ARMATURE, FIRST REMOVE THE ARMATURE FROM THE CHASSIS, THEN INSTALL T-NUT REPAIR INSERT.
6. IN THE CASE THAT THE OLD HOLE CAN NOT BE REUSED, THE T-NUT REPAIR INSERT CAN BE INSTALLED IN ANY PLACE. BUT MAINTAIN CLOSE PROXIMITY TO THE OLD LOCATION. IN THIS CASE DRILL ADDITIONAL 5/16' (8mm) CENTER HOLE TO ACCOMMODATTE T-NUT INSERT BARREL.

96-104133 Rev. 1.0