Internal cleaning of HYDRIM Backflow Preventer

Product Support 

Description:  

Internal cleaning of Backflow Preventer. This procedure applies to All HYDRIM® G4 WD units C61WD & M2 G4 for all markets except North America.

The HYDRIM C61w G4 is equipped with an air gap / anti-suction device to prevent
backflow of dirty water into the water supply. No other air gap device is necessary.

NOTE: When a cycle is started (P1, P2, P3) the air gap is filled four times before the circulation pump starts.

Referenced Parts

Backflow Preventer 01-113318S (C61WD) and 01-113853S (M2 G4).

Required Materials

The following additional materials are required for this step:

  • 1x large syringe with tube  (or sprayer bottle)
  • 1x citric acid-based instrument neutralizer (for example, Dr. Weigert Neutralizer or distilled white vinegar)
  • A long flexible bottle brush may be helpful

Cleaning Procedure: Backflow Preventer/Air Gap

  1. Remove the top cover from the HYDRIM unit and the left side panel (C61WD) or rear panel (M2 G4).
  2. Fill a large syringe with 50ml neutralizer and then inject it through the opening on the Air Gap (Fig. 1).
  3. Turn on the device and go into the technician menu: Technician menu / Diagnostic tools / Component test / Filling Draining (Fig. 2).
  4. IMPORTANT: Start the filling process by pressing the “On” button. Once the unit reaches the position “E = On, F=On,” immediately press the “On” button again. This action will prevent the unit from activating the Air Gap pump and emptying the air gap.
  • WARNING: The device may still contain residues of dirty water. Be sure to drain them via the drain hose as described, and do not pump into the wash chamber by starting the Air Gap pump. Otherwise, these contaminants can enter the pump and damage it.
  1. Leave the liquid to work for approximately 10 minutes. Tap gently on the side of the Backflow Preventer with the palm of your hand so that the agent spreads and the dirt dissolves.
  2. Drain the liquid via the drain hose (at the bottom of the unit). Repeat the process with the filling from point 2 on, for further visible contamination (e.g., shadows, coverings, biofilm).
  3. When no more contamination is visible, apply another 25ml neutralizer with the syringe and inject it into the air gap via the opening. Fill the Backflow Preventer again by going into the Technician Menu / Diagnostics Tools / Component Testing / Filling Draining. This time, do not press the “On” button again after reaching the F-switch position. This allows the Air Gap pump to start, which cleans and descales it as it empties the air gap of the solution. Repeat this step if necessary.
Figure 1 and 2

  • Timeout on filling/emptying up the Air Gap reservoir/chamber. C61 – 150s. For RO – 120s. L110 – 200s. For RO – 600s. M2 – 180s. For RO – 200s

Probable Cause

  • Water supply issue
  • Water valves failure
  • Air gap water pump failure
  • Air gap valve failure
  • Air gap Full/Empty level switches failure
  • Chamber water level pressure switch malfunction
  • Overflow switch malfunction triggering evacuation pump

Tech Troubleshoot

  • Water supply issue :water valves, water pump or air gap valve have failed.
  • Air gap Full/Empty level switches failed.
  • Chamber water level pressure switch malfunction.
  • Overflow switch malfunction triggering evacuation pump.

User Troubleshoot

  • Contact dealer service department.

Sources:


© 2019 SciCan Ltd., SciCan is a COLTENE Group Company. All rights reserved.

www.scican.com SciCan Ltd., 1440 Don Mills Road, Toronto, ON M3B 3P9, Canada 

Ensure Autoclave Monitoring

SKU SCI-BISK-120
$413.50
Preorder
Ensure
Out of stock, available for pre-order
1
Save this product for later
Share this product with your friends
Have questions?
Ensure Autoclave Monitoring
Product Details
Brand: SciCan
Model: Ensure
UPC: SCI-BISK-120
Weight: 3 lbs

Think of it as a routine checkup for your sterilizer.

The ENSURE line of Infection Control monitoring products ensures your equipment is performing at its peak. It includes a range of products that are simple, easy-to-use, and help to ensure your clients are being kept safe with every cycle.

The Incubator

The unique incubator allows you to monitor up to 5 sterilizers at one time, providing you with easy to read results that show whether sterility has been achieved... or not. The ENSURE system can satisfy even the most stringent international regulatory requirements.

No need to wonder, when it’s this easy to know. The ENSURE Incubator, a 10-well incubator, is a very unique unit unlike any in the market today. With an integrated vial crusher and built in timer for up to five different programs, the SciCan ENSURE BI system leaves no room for error in time.

Biological Indicators (BI) "Spore Tests"

The ENSURE BI’s are fast, reliable, self-contained indicators for results in as little as 24 hours. BI’s are the only cost effective method to ensure effective performance of your sterilizer.

The ENSURE BI’s are available in a starter kit system that consist of 25 BI’s, the unique incubator and an easy to fill-in log book. The starter kit is the most cost effective, accurate system for monitoring your steam autoclave directly in your office in only 24 hours.

Specification

Part Number
Description
SCI-134 Class 4 ENSURE CI strips 132-134°C (250 count)
SCI-BI25 ENSURE BI's (25 count refill box)
SCI-BI100 ENSURE BI's (100 count refill box)
SCI-BISK-120 ENSURE system, Starter kit (25 BI's, incubator, log book)
SCI-BINC-120 ENSURE incubator (incubator only)
SCI-BILOG ENSURE BI monitoring system log book (PDF Download)

Ensure SDS

Customer reviews on “Ensure Autoclave Monitoring”
Reviews only from verified customers
No reviews yet. You can buy this product and be the first to leave a review.