Stat-Dri Plus

STAT-DRI Plus Drying / Rinse Agent For faster drying

For faster drying and seal lubrication coat the STATIM Cassette: After cleaning and drying the cassette, treat the interior surfaces with STAT-DRI Plus.
Add to the rinse aid dispenser on your HYDRIM instrument washer for effective drying results. Use StatDri Plus Drying agent for dramatically improved STATIM, BRAVO and HYDRIM drying and to extend the life of your STATIM seal.

You’ll notice the difference on your first cycle!

Stat-Dri Use Protocol

STATIM 2000/5000 Operator’s Manual 95-108027 Rev 5.0.
Copyright 2008 SciCan Ltd . All rights reserved.

3.9 Preparing Unit for Use

Once the unit is installed and before any instruments are sterilized, run two Wrapped Cycles (see Section 4.8 Statim 2000 – Running a Cycle and 4.10 Statim 5000 – Running a Cycle). Remove the cassette once it has cooled. Clean the top (lid) and bottom (tray) sections using a soft cloth to wipe the inside surfaces and then rinse thoroughly with tap water. Once the cassette is clean and dry, coat the inside surfaces with STAT-DRI.

STAT-DRI

Treatment of the interior surfaces of the cassette with the STAT-DRI drying agent, provided with your unit, will enhance the drying process. (Replacement bottles are available from SciCan, order number 2OZPLUS, 8OZPLUST, 32OZPLUS).

4.3 Statim 5000 — STAT-DRI Plates

The Statim 5000 cassette uses STAT-DRI plates to enhance the drying process for wrapped instruments. The adjustable plates are designed for the unwrapped instruments rack and eliminate the need for a separate wrapped instrument rack. Up to ten plates can be arranged along the length of the rack. Each unit is shipped with five STAT-DRI plates. Additional plates may be ordered from SciCan (SciCan part number 01-103935).

To install and adjust a STAT-DRI plate, follow these steps:

  1. Hold a plate over the desired position, tilted slightly forward with the tabs down.
  2. Insert the tabs between the mesh of the unwrapped instrument rack.
  3. Each tab has an elongated slot. Slide the plate until the wire is at the end of each slot.
  4. Release the plate to the resting position.
  5. Prepare and load instruments for sterilization.

5.1 Cleaning the Cassette

Keeping the Statim cassette clean is good clinical practice and assists in the function of the unit. SciCan recommends that the interior surface be cleaned at least once a week. Use dishwashing soap or a mild detergent that does not contain chlorine. Scrub the inside of the cassette with a cleaning pad designed for use with Teflon™ coated surfaces. After scouring, rinse thoroughly with water to remove all traces of the detergent. Cleaning the inside of your cassette is very important if you regularly sterilize lubricated instruments. Coating the entire inside surface with STAT-DRI / STAT-DRI PLUS drying agent induces water to form an even coat on the inside surface, without beading. The water in contact with the hot cassette surfaces also evaporates much more efficiently. Spotting is minimized and instruments dry much better. STAT-DRI / STAT-DRI PLUS should be applied every 10 cycles, and after every cassette cleaning.



Stat-Dri vs. Jet-Dry

Stat-Dri Plus is specially formulated to protect stainless steel instruments. All microbiological sterilization efficacy testing has been done using Stat-Dri. We have not conducted such tests using Jet-Dry, and therefore cannot provide any data to support using any Jet-Dry products as an alternative.

Keep in mind there are several flavours of Jet-Dry on the market, including Jet-Dry with baking soda, Jet-Dry Turbo-Dry, etc. all of which have slightly different chemical composition. As a result, customers who choose to use Jet-Dry instead of Stat-Dri do so at their ‘own risk’.

Courtesy of Statim.us

 

StatClave G4

G4-621103
$9,150
STATCLAVE G4 is a Class B vacuum sterilizer with an 11 inch chamber as a companion to your STATIM for loads of packages
In stock
1
Product Details
Brand: SciCan
Model: StatClave G4

*(120V) Wifi enabled, waste water direct to drain included


This changes everything. Again.

Introducing the STATCLAVE™ Chamber Autoclave


With the reliability and trust you’ve come to expect from STATIM, the STATCLAVE G4 is designed to deliver speed, capacity and efficiency to your practice. STATCLAVE is a Class B vacuum sterilizer that incorporates both pre-sterilization vacuum to improve steam penetration and post-sterilization vacuum to improve drying. Its 11” chamber and vacuum-assisted closed-door drying efficiently delivers sterilized and dried wrapped instruments.


Fast

STAT*CLAVE* G4

STATCLAVE can efficiently sterilize and dry wrapped instrument loads in as little as 38 minutes.

Dry

Closed-door drying and vacuum technology results in perfectly dried instruments every time.

G4 Technology

SciCan’s G4 equipment can connect and send information to your smart devices or computer so you can have cycle data, instructions, and maintenance notifications at your fingertips.

Capacity

STATCLAVE’s 11” chamber can sterilize up to four full-size cassettes and 4 exam cassettes or 20 pouches with included pouch racks.

Versatile

STATCLAVE offers a number of filling & draining options, including easily accessible and cleanable top reservoirs, front manual fill and a rear automatic fill option that can be connected to the VistaPure water filtration system for automated filling.

Gentle

Fresh water is used in every cycle, reducing the buildup of oils and debris.


STATCLAVE G4 Chamber Autoclave

Featuring the STATCLAVE G4 Chamber Autoclave for sterilizing dental instruments.


STATCLAVE G4 Launch at PDC 2019

Ted Gray and Melissa Summerfield discuss the latest innovation from SciCan, the STATCLAVE G4 Chamber Autoclave.



STATCLAVE G4 In-Service

An overview of installing, operating, and maintaining the SciCan STATCLAVE G4 Chamber Autoclave.

Cycle Times and Maximum Load Weights (120v)

Cycles
Temperature / Time
Cycle Times*
Maximum Load Weight
STATCLAVE G4
Solid Unwrapped
270°F (132°C) / 4 mins
18 mins
13.2 lbs (6 kg)
Solid Wrapped
270°F (132°C) / 4 mins
44 mins
13.2 lbs (6 kg)
Hollow Unwrapped
270°F (132°C) / 4 mins
23 mins
13.2 lbs (6 kg)
Hollow Wrapped
270°F (132°C) / 4 mins
50 mins
13.2 lbs (6 kg)
Textiles/Porous
270°F (132°C) / 4 mins
49 mins
4.4 lbs (2 kg)
Rubber & Plastic
250°F (121°C) / 20 mins
36 mins
4.4 lbs (2 kg)

*Cycle time will vary depending on instrument load.



Unit Specifications

STATCLAVE G4
Unit size (L x W x H)†
25” x 17.75" x 19.5" / 635 x 450 x 495mm
Volume
6.87 gal (26 L)
Chamber Dimensions (D x L)
11" diameter, 15" depth / 297 x 381mm
Weight
61.6 kg (135.8 lbs) (without water and packaging)
Voltage
120V, 60Hz, 12 A

†On the right side, allow 2” (50mm) of space for ventilation. Allow for at least 7” (180mm) above unit to enable access to reservoirs. If less than 7” (180mm), screw on hinges can be removed to allow reservoir lid to slide in and out of position.