STAT-DRI Plus Drying / Rinse Agent For faster drying

For faster drying and seal lubrication coat the STATIM Cassette: After cleaning and drying the cassette, treat the interior surfaces with STAT-DRI Plus.
Add to the rinse aid dispenser on your HYDRIM instrument washer for effective drying results. Use StatDri Plus Drying agent for dramatically improved STATIM, BRAVO and HYDRIM drying and to extend the life of your STATIM seal.

You’ll notice the difference on your first cycle!

Stat-Dri Use Protocol

STATIM 2000/5000 Operator’s Manual 95-108027 Rev 5.0.
Copyright 2008 SciCan Ltd . All rights reserved.

3.9 Preparing Unit for Use

Once the unit is installed and before any instruments are sterilized, run two Wrapped Cycles (see Section 4.8 Statim 2000 – Running a Cycle and 4.10 Statim 5000 – Running a Cycle). Remove the cassette once it has cooled. Clean the top (lid) and bottom (tray) sections using a soft cloth to wipe the inside surfaces and then rinse thoroughly with tap water. Once the cassette is clean and dry, coat the inside surfaces with STAT-DRI.

STAT-DRI

Treatment of the interior surfaces of the cassette with the STAT-DRI drying agent, provided with your unit, will enhance the drying process. (Replacement bottles are available from SciCan, order number 2OZPLUS, 8OZPLUST, 32OZPLUS).

4.3 Statim 5000 — STAT-DRI Plates

The Statim 5000 cassette uses STAT-DRI plates to enhance the drying process for wrapped instruments. The adjustable plates are designed for the unwrapped instruments rack and eliminate the need for a separate wrapped instrument rack. Up to ten plates can be arranged along the length of the rack. Each unit is shipped with five STAT-DRI plates. Additional plates may be ordered from SciCan (SciCan part number 01-103935).

To install and adjust a STAT-DRI plate, follow these steps:

  1. Hold a plate over the desired position, tilted slightly forward with the tabs down.
  2. Insert the tabs between the mesh of the unwrapped instrument rack.
  3. Each tab has an elongated slot. Slide the plate until the wire is at the end of each slot.
  4. Release the plate to the resting position.
  5. Prepare and load instruments for sterilization.

5.1 Cleaning the Cassette

Keeping the Statim cassette clean is good clinical practice and assists in the function of the unit. SciCan recommends that the interior surface be cleaned at least once a week. Use dishwashing soap or a mild detergent that does not contain chlorine. Scrub the inside of the cassette with a cleaning pad designed for use with Teflon™ coated surfaces. After scouring, rinse thoroughly with water to remove all traces of the detergent. Cleaning the inside of your cassette is very important if you regularly sterilize lubricated instruments. Coating the entire inside surface with STAT-DRI / STAT-DRI PLUS drying agent induces water to form an even coat on the inside surface, without beading. The water in contact with the hot cassette surfaces also evaporates much more efficiently. Spotting is minimized and instruments dry much better. STAT-DRI / STAT-DRI PLUS should be applied every 10 cycles, and after every cassette cleaning.



Stat-Dri vs. Jet-Dry

Stat-Dri Plus is specially formulated to protect stainless steel instruments. All microbiological sterilization efficacy testing has been done using Stat-Dri. We have not conducted such tests using Jet-Dry, and therefore cannot provide any data to support using any Jet-Dry products as an alternative.

Keep in mind there are several flavours of Jet-Dry on the market, including Jet-Dry with baking soda, Jet-Dry Turbo-Dry, etc. all of which have slightly different chemical composition. As a result, customers who choose to use Jet-Dry instead of Stat-Dri do so at their ‘own risk’.

Courtesy of Statim.us

 

Bravo G4 Chamber Autoclave

7A620010
$8,495
In stock
1
Product Details
Brand: SciCan
Model: Bravo 21V

The BRAVO G4 Chamber Autoclave is an EN13060 compliant Class B sterilizer that addresses a dental office’s need for speed, capacity, safety, and efficiency. Its cutting-edge sterilization technology incorporates both pre-sterilization vacuum to improve steam penetration and post-sterilization vacuum to improve drying.

Efficient

  • The BRAVO G4 chamber autoclaves offer efficient cycle times, sterilizing and drying instruments in as little as 33 minutes.
  • The delayed start feature allows the user to program cycles to begin at a later time; an effective feature when programming test cycles to begin before the start of the day.

Versatile

  • Offering enough capacity to meet the needs of any size practice, the BRAVO G4 comes in 17L, 22L, or 28L chamber size.
  • The BRAVO G4 offers numerous filling and draining options, including manual top and front filling, and automatic filling and draining through ports at the back.
  • The accessory water equipment and the BRAVO G4 footprint allows options for any Stericentre set-up.

G4 Technology

Connectivity & Data Logging

  • Integrated G4 technology in the BRAVO G4 allows for reliable load release, documentation and data storage.
  • WiFi enabled G4 technology can connect and send cycle data to your smart devices or computer, providing easily accessible cycle data and maintenance information.
  • To help optimise office processes, the BRAVO G4 can be configured to email cycle information or error messages directly to office staff and/or service technicians.
Cycle 17L 22L 28L
Cycle Type Cycle Time (mins) Max. Load (kg) Cycle Time (mins) Max. Load (kg) Cycle Time (mins) Max. Load (kg)
134°C Universal B 42 6 46 7.5 56 9
134°C Hollow Unwrapped S 35 6 39 7.5 44 9
134°C Solid Wrapped S 33 6 39 7.5 45 9
134°C Prion B 56 6 60 7.5 70 9
121°C Universal B 58 6 63 7.5 69 9
134°C Helix Test N/A 20 N/A 24 N/A 24 N/A
Vacuum Test N/A 18 N/A 18 N/A 18 N/A
Custom (134°C or 121°C) N/A N/A N/A N/A N/A N/A N/A


BRAVO G4 Sterilization, The Smarter Way.

The BRAVO G4, a Type B sterilizer, incorporates both pre-sterilization vacuum to improve steam penetration and post-sterilization vacuum to improve drying. The option of 17L, 22L or 28L units and efficient turnover times ensure clinicians always have an ample supply of instruments ready for use, regardless of practice size