STATIM Maintenance

General StatIM Preventative Maintenance:

To ensure trouble-free performance, both the operator and the dealer must follow a preventative maintenance schedule.

NOTE: Please refer to your National, Regional, State or Safety laws for any additional reoccurring user testing that may be required.

OperatorPreventative maintenance actionPart

Daily

Water Reservoir💧Replace water as needed. (Use distilled water <5ppm TDS)
💧Empty the waste bottle every time you refill the water reservoir.
💧For ophthalmic use, drain at the end of every workday, leave empty, and refill at the start of the next workday.
Water distiller
VistaPure
TDS Meter
Waste Bottle💧Empty the waste bottle every time you refill the water reservoir.
💧Fill the bottle with water, up to the MIN line marking.
Waste bottle and parts

Weekly

Cassette💧Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine.
💧Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces.

💧After removing all traces of the detergent, treat interior surfaces of the cassette with the STAT-DRI™Plus drying agent to enhance the drying process.

Cassette
STAT-DRI™ Plus
Biological and/or Air Filter🔎Check the filter for dirt and moisture.
🔁Replace if dirty. ⚠️Call for service if wet.💧
Biological Filter
Air Filter
Water Filter🔎 Check the water reservoir filter every week and clean if necessary. Replace only if necessary.Reservoir filter

Every 6 months

Cassette Seal🔁Replace every 500 cycles or six months (whichever is first), or whenever necessary.
🔎Video: How to change the cassette seal
Cassette seal
Biological and/or Air Filter🔁 Replace every 500 cycles or six months (whichever is first).Maintenance Kit:
Cassette seal and both air filters
Regular preventative maintenance

TechnicianOnce a yearPart
Cassette🔎Check the tray, lid and seal for damage. Replace if necessary.
Biological Filter🔎Inspect the biological filter for moisture.
Solenoid Valve🔎Inspect the valve and clean if dirty. Replace the plunger if defective.
Pump🔎Replace in-line water filter.
Check Valve🔎Remove the exhaust tube from the back of the unit during the air drying phase. 🔎Check for air coming from the fitting.

 🔎Remove the air compressor tube from the check valve inlet while running a cycle. Make sure no steam is leaking from the valve. Replace if there are any leaks

Water Reservoir🔎Check the reservoir for dirt. Clean and rinse with steam process distilled water if necessary.
Calibration🔎Calibrate the unit.
Annual preventative maintenance

 

More information about preventative maintenance subscriptions

Here is the free poster and links to individual parts to keep on hand to keep things working smoothly:

Consider also a second sterilization cassette: STATIM 2000 Sterilization Cassette [parts available separately]

Also, to protect the cover of your Statim: STATMAT protective cover for STATIM 2000 + G4


More details:

reservoir_cap_filter

Water Reservoir

Replace water as needed with pure distilled water (Less than 5 parts per million of Total Dissolved Solids)

For ophthalmic use, drain at the end of every workday, leave empty, and refill at the start of the next workday.

Statim reservoir view of exit hole, float and water quality sensor

🔎Statim reservoir view of exit hole, float and water quality sensor

Waste Bottle complete with tubing is included with the STATIM sterilizer

Waste Bottle

Empty the waste bottle every time you refill the water reservoir.

Fill the bottle with water, up to the MIN line marking.

Get a new waste bottle or parts, such as tubing or the top fixture.

cassette_volumes

Cassette

Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine.

Scrub the inside with a cleaning pad designed for use with Teflon®-coated surfaces.

STAT-DRI Plus dramatically shortens drying time in your autoclave!

STAT-DRI Plus dramatically shortens drying time in your autoclave!

STAT-DRI Plus refills can be ordered in our shop

Water Reservoir Filter and Cap Kit

STATIM Water Reservoir Filter and Cap Kit

Check the water reservoir filter every week and clean if necessary.

Replace Filter only if necessary.

IMPORTANT UPGRADE: Get a new cap and water filter kit if your STATIM does not have one.

Replacing the water reservoir filter: read full article

Air and Biological Air Filters for STATIM 2000 (shown)

Biological Filter

Air and Biological Air Filters for STATIM 2000 (compressor shown open)

Replace both air filters every 500 cycles or six months (whichever is first).

Get a new biological filter. Older units will not have the biological air filter.

⚠️ Call for service if wet. The check valve will need to be replaced.

This biological filter is wet. 

The check valve must be replaced

⚠️This biological filter is wet.
The check valve must be replaced

Replace your STATIM air filters every six months or else it will look like the gross dirty yellow one top left! (STATIM 2000 air compressor filter shown)

Air Filter

Replace both your STATIM air filters every six months or else it will look like the dirty yellow one top left!
(STATIM 2000 air compressor filter shown)

Replace every 500 cycles or six months (whichever is first). Get a new air filter.

Seal & Lube Kits

STATIM Cassette Seal & Lube Kits

[Seals go green! Read full article]

Replace every 500 cycles or six months (whichever is first), or whenever necessary. Get a new cassette seal.

If you need further assistance, a STATIM user’s manual, Reference Poster, or a Troubleshooting Guide, please contact us

Brian Skellie and our team will help!

Statim 5000 G4

G4-201103
$7,380
Larger volume for wrapped loads, same Statim technology!
Pre-order
1
Product Details
Brand: SciCan
Model: 5000 G4

Introducing The New Generation of Statim, The Statim G4 Series

Download brochure

Canonsburg, PA – SciCan has set the benchmark for sterilization and instrument reprocessing technology.

For more than 20 years SciCan has been developing and manufacturing highly innovative infection control products that address the need for versatility, flexibility, and efficiency in today’s environments.

SciCan is proud to unveil the newest STATIM family member, the G4 series. The STATIM G4 series is the same renowned and trusted autoclave it has been for over 20 years but now boasts a new contemporary look, functionality and connectivity that is the first of its kind. The G4 technology will change the way you interact by providing a direct channel of communication through the Internet to increase productivity and limit downtime.

Still powered by SciCan’s signature steam technology to provide sterilization at speeds faster than conventional chambered autoclaves, the STATIM G4 series has been drastically upgraded with a level of interactivity not seen before.

New look

The STATIM G4 series has been modernized with a new fascia that is fresh and contemporary.

Color touch screen

A large 3.5” high-resolution touchscreen offers a vivid display of messages and current cycle information all with extraordinary clarity. The bubble level has been integrated into the high-resolution touchscreen.

Built-in data logger

Gone are the days of messy thermal paper, clumsy printers or potential failures with external data loggers. With increasing stringency amongst national guidelines, the STATIM G4 series now saves all the data from each cycle. This data can be copied onto the included USB drive, viewed on screen, or remotely stored in the Cloud via email.*

STATIM G4 technology

SciCan's STATIM G4 series Technology offers a platform with endless possibilities. The product expansion and modes of communication will provide visibility from every facet, from usability to troubleshooting.

This is a very exciting time for SciCan, as we are embarking into new territory with cutting edge technology. SciCan has always demonstrated innovation and this is just another example of how we are setting new standards for infection control.

*Based on your email providers storage capabilities


This model has all the benefits of our classic STATIM 5000 Cassette Autoclave

The STATIM 5000 "Classic" Cassette Autoclave has been superseded by the G4 model

  • Ultra-fast 9 minute unwrapped cycle.
  • 17.5 minute wrapped cycle
  • Exceptional instrument turnover
  • 15 minute Dri-TecR Drying Technology
  • Reduced instrument investment
  • Prolongs instrument life

Larger than the 2000, the Statim 5000 has an additional sterilization cycle and a roomier cassette designed to process bulkier instruments and greater loads. It's a reliable cassette sterilizer that's compact enough to fit in patient treatment areas and fast enough to sterilize expensive Lasik, PRK, endoscopic and cataract instruments, including phaco handpieces, between procedures - it can complete an unwrapped sterilization cycle in just nine minutes and a wrapped cycle in just 17.5 minutes.*

The Statim 5000 is fully automatic. Simply select one of four sterilization programs and with a single touch of the keypad, begin the cycle. An advanced microprocessor controls all functions and maintains optimal sterilization conditions, from start to finish.

SciCan's Statim cassette sterilizers have a biological effectiveness that has been proven in tests for hollow and solid instruments at many internationally recognized institutes in Canada, the United States and Europe.

* Cycle times do not include drying.


STATIM FAQs: Click here

ACCESSORIES: Get STATIM Accessories here

SPECIFICATIONS:

Unit Size : L 555mm x W 415mm x H 190mm
L 21.75" x W 16.25" x H 7.5"
Cassette Internal Dimensions : L 380mm x W 180mm x H 75mm
L 15" x W 7" x H 3"
Reservoir Capacity : 4.0 litres (distilled water)
Weight without water : 73lbs or 33kg
Power Consumption : 110V, 60Hz, 11A