Save with OPTIM OS1 | United States

save with optim os1

Stop throwing away money on wipes

Tell us about your studio and we’ll show you how much you can save.

OPTIM OS1

Here is a calculator for how much you can save with OPTIM OS1 wipes

Save Money

Save Money

Offices using Quats-alcohol-based wipes that switch to OPTIM OS1 can save over $2,000 per year on average.

Shorten Turn Over Time

Shorten Turn Over Time

A hospital-grade cleaner-disinfectant with 60-second kill times for the toughest viruses, bacteria, and yeast, and 45 seconds for TB.

Reduce Wipe Waste

Reduce Wipe Waste

Wipe down the same surface area that would take two competitor wipes to complete effectively.

How they tested it.

In a recent Dental Advisor study² on surface disinfectant wipes, OPTIM OS1 outperformed the other brands, including a leading quats-alcohol wipe. Not only did it remain effective for its full 1-minute contact time, but it also covered a much larger surface area than the competitor wipes.

To test the wipes, researchers used a 12.5-square-foot table divided into four sections. Each wipe was used to wipe down one quarter section. Researchers then waited for the kill time to elapse before moving to the next section with the same wipe. OPTIM OS1 was able to effectively clean and disinfect two times more surface area than the leading quats-alcohol-based wipe.

View Dental Advisor study below

Dental Study

Pump 110V Statim 2000

SKU 01-100568S
$293
Pump type
In stock
1
Save this product for later
Have questions?
Share this product with your friends
Customer reviews
Reviews only from verified customers
No reviews yet. You can buy this product and be the first to leave a review.
Pump 110V Statim 2000
Product Details
Brand: SciCan
Weight: 2 lbs

Water pump. For STATIM 1000 & 2000 units, 120V.

Water Pumps

STATIM units have been manufactured using three different pumps. Before servicing, you must first determine which model of pump the unit contains (See Figure 1). Important Note: Always make sure that there is sufficient steam-process distilled water in the reservoir prior to testing STATIM pumps.

Identifying Pump Types

Fluid-O-Tech Pump

The Fluid-O-Tech pump is generally factory fitted to older units with revision 3.x/4/x controller boards up to approximately 1994-1995. It is a small rectangular pump with a diode plug attached to the top. The pump rests on a vibration-damping pad on the bottom of the chassis and is held in place with a wire bracket. The pump is secured to the bracket by a cable tie at each end. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

SciCan Pump

The SciCan pump is generally factory fitted to units with revision 2.x/5.x/6.x and some early 7.x controller boards from 1995 up to mid 2008. It has a similar footprint to the Fluid-O-Tech pump, however the diode is now integrated into the pump. The pump is mounted on two vibration damping rubber brackets. Each bracket is secured to the chassis with two screws. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

Ulka Pump

The Ulka pump is factory fitted to units with revision 7.x controller boards only from mid 2008 onwards. It is slightly smaller than the SciCan pump and can be identified by the configuration of the wires on top of the pump. The two visible connections are located adjacent and parallel to each other whereas the two visible connections on the SciCan pump are opposing each other and have a thermal switch connecting them. The pump is mounted on two vibration damping rubber brackets. Each bracket is secured to the chassis with two screws. The black lead is attached to controller board J1-5 (LINE) and the white lead is attached to controller board J1-6 (NEUTRAL).

Note: Important information on pump interchangeability:

  • The Fluid-O-Tech pump is no longer available, but units with this pump can be retrofitted with the SciCan pump ONLY.
  • The SciCan pump and Ulka pump are NOT interchangeable. They are both still available. If replacing the pump during maintenance or servicing, the same pump must be fitted to ensure the unit works correctly.