Introducing SciCan OPTIM OS1 Liquid

Rapid contact times for best results

  • 60-second kill times for viruses, bacteria, and yeast
  • Kills TB Bacteria in only 45 seconds and Candida auris in 60 seconds

Safe for use on hard, non-porous clinical surfaces

  • Cleans and disinfects stainless steel, anodized aluminum, non-porous upholstery, Corian countertops, silicone, and more

Formulated with the user in mind

  • VOC-free, fragrance-free and dye-free

Emerging Pathogen Claim ensures your clinic is ready for the next viral
threat

  • Qualifies for EPA’s Emerging Viral Pathogen Claim
  • Kills SARS-CoV-2, the virus that causes COVID-19, in only 30 seconds

You can learn more about OPTIM’s newest formula below

> Product Brochure EN

> Safety Data Sheet (Wipes) EN

> Safety Data Sheet (Liquid) EN

> Manual Instructions for Use EN

OPTIM OS1 Wipes Quick Reference Guide

OPTIM OS1 SOLUTION Liquid Spray

Spray Ready-to-Use OPTIM OS1 liquid onto a SciCan Dry Towelette. Do not directly spray onto the surface.

Spray Ready-to-Use OPTIM OS1 liquid onto a SciCan Dry Towelette. Do not directly spray onto the surface.


If surface is not heavily soiled, clean and disinfect surface with one wipe and allow surface to remain wet for the contact time indicated on the label.

If surface is visibly soiled, clean surface with one
towelette and discard. Use a second towelette to disinfect and allow surface to remain wet for the contact time indicated on the label.

OPTIM OS1 Ready-to-Use Wipes

Pull out OPTIM OS1 towelette at a 45 degree angle. Close lid.

If surface is not heavily soiled, clean and disinfect surface with one wipe and allow surface to remain wet for the contact time indicated on the label.

If surface is visibly soiled, clean surface with one towelette and discard. Use a second towelette to disinfect and allow surface to remain wet for the contact time indicated on the label.

Procedure for disinfecting PVS elastomeric impression materials with OPTIM OS1

  1. Once the impression is taken, rinse it well under running tap water to remove blood and saliva.
  2. After rinsing, gently shake the impression within the sink basin to remove any adherent water with minimal splatter.
  3. Spray the entire impression with OPTIM OS1. Use the contact times as stated on the label.
  4. Rinse the impression thoroughly with tap water to remove any excess disinfectant. After a thorough rinse, gently
    shake the impression inside the sink basin to remove any adherent water with minimal splatter.
  5. Once dry, place the disinfected impression in a sealed bag.
    The impression is now safe and ready for further processing.
PVS elastomeric impression materials

Switching to OPTIM OS1

  • Give all surfaces a good cleaning with soap and water before switching to OPTIM OS1.
  • After switching to OPTIM OS1, surfaces should to be periodically rinsed with water (recommended no less than once per week).
  • Common practice for all chemical disinfectants is to periodically rinse and dry surfaces to maintain the surface integrity.
  • Avoid mixing of chemistries.

Annual maintenance Parts for the Statim 2000 classic and G4

Keep it clean & functional!

Here is the free poster and links to individual parts to keep on hand to keep things working smoothly:
Quick Reference Poster Maintenance STATIM 2000 & 5000
  1. STATIM 2000 Cassette Seals
  2. Air compressor filter [foam donut] for STATIM 2000
  3. Biological Air compressor filter [external clear plastic] for STATIM 2000, 5000 & 7000
  4. Solenoid Plunger Repair Kit
  5. Check valve replacement kit 2000
  6. 15A Fuses for STATIM 2000
  7. Water Filter In-Line Kit
  8. STATDRI Plus Drying / Rinse Agent
Consider also a second sterilization cassette:
STATIM 2000 Sterilization Cassette [parts available separately]
Also, to protect the cover of your Statim:
STATMAT protective cover for STATIM 2000 + G4

Time for Preventative Maintenance!

Cassette

• Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine.
• Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces.
• After removing all traces of the detergent, treat interior surfaces of the cassette with the STAT-DRI Plus drying agent to enhance the drying process.

Biological and/or Air Filter

• Check the filter for dirt and moisture. Replace if dirty. Call for service if wet.

Water Filter

• Check the water reservoir filter every week and clean if necessary. Replace only if necessary.

Every 6 months

Cassette Seal • Replace every 500 cycles or six months (whichever is first), or whenever necessary.
Biological and/or Air Filter • Replace every 500 cycles or six months (whichever is first).

https://statim.us/info/maintenance

Refine by
 

Filters

Filters
Availability
 
Availability
Price
 
Price
from
to
$0
$154
Change every 6 months
Air compressor filter [foam donut] for STATIM 2000
Air compressor filter [foam donut] for STATIM 2000
$22.66
Air compressor filter [foam donut] for STATIM 2000
This air filter, behind the circular metal plate on STATIM 2000 and G4 units, must be changed every six months