OPTIM 33TB cleaning protocol

SciCan_OPTIM_USA

OPTIM™33 TB Quick Reference Guide

SWITCHING TO OPTIM 33 TB

Give all surfaces a rinse with mild soap & water before switching to OPTIM.

SciCan_OPTIM_USAQUATs (Quaternary Ammonium Compounds) or phenol based disinfectants are common active ingredients in the dental operatory today and are known to leave residual active chemistry on surfaces. AHP (Accelerated Hydrogen Peroxide) is an oxidizer and will remove chemical residues left by other surface disinfectants. There may be odor or film development if surfaces are not rinsed (damp wiped) properly with soap & water. This odor or film should however dissipate on average within 7 to 10 days of continual use as AHP will remove the residual active ingredient.

Surfaces need to be periodically rinsed with water!

Dissents or complaints happen when users are not informed on the correct protocol when using the product. It is good practice to periodically (at least once a week) rinse or wipe surfaces with a damp cloth to remove surfactants (detergents) or residual active ingredients that can build up over time with the use of ANY disinfectant cleaner.

No mixing of chemistries!

Mixing of chemistries in operatory i.e. using quats-alcohol/phenol disinfectant and OPTIM 33 TB in the same operatory or refilling OPTIM into chemical container used previously for other chemicals is not recommended. It is best practice to never mix chemistries (any kind) no matter what industry you are in.

On which surfaces can OPTIM 33 TB be used?

Non-porous hand contact environmental surfaces including Naugahyde (vinyl coated upholstery fabric).

NOTE: Soft surfaces such as chairs and cables should be periodically rinsed with water or wiped with a damp cloth to prevent the build-up of disinfectant or surfactants. As best practice, for any kind of disinfectant or cleaners used, chairs should be wiped with a damp cloth at the end of the day or at least once a week.

PRECAUTIONS:

– Surfaces that are composed of brass or copper, or other ferrous metals may show signs of discoloration or pitting with prolonged exposure.

– Anodized aluminum (often used on hand-pieces) and carbon tipped instruments should be avoided (i.e. burrs). Material impact can be reduced with rinsing or damp wiping.

We advise against wrapping gauze pad or ‘sleeves’ soaked in disinfectant on hand-pieces.


OPTIM 33 TB Solution

OPTIM 33TBSpray – Wipe – Spray Protocol

Visibly Soiled Surface – Clean before disinfecting

  1. Spray OPTIM 33 TB directly onto surface.
  2. Wipe visible soil with Stat-Wipes or clean paper towel to clean surface.
  3. Discard soiled towelette.
  4. Continue to disinfection step

Visibly Clean Surface – Disinfect pre-clean surface

  1. Spray OPTIM 33TB to cover the pre-clean surface. Allow the disinfectant to remain
  2. undisturbed on the surface for the contact time indicated on the label.
  3. *Do not overspray as this can create pooling of liquid. OPTIM 33TB is water based not alcohol based.
  4. Wipe with Stat-Wipes or clean paper towel.
  5. Discard used towelette.

*Using gauze dipped in solutions for wiping can cause pooling of liquid on surface. Wipe surface dry with paper towel if surface is still wet after desired contact time is achieved.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One- Step Cleaner Disinfectant. Therefore, OPTIM can be used on visibly clean surface. Depending on the soil load on the surface, the 2 step process is not required due to this EPA claim.


OPTIM 33 TB Wipes

Standard Protocol for Visibly Soiled Surface

Clean before disinfecting

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Clean surface with towelette to remove organic matter or visible blood.
  3. Discard used towellete.

Disinfect After Cleaning

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Disinfect surface with towelette.
  3. Allow surface to remain wet for the recommended contact time.
  4. Discard used towelette.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One-Step Cleaner Disinfectant. Depending on the soil load on the surface, users have the choice of cleaning the surface with one side of the wipe and disinfecting the surface with the reverse side of the wipes. Always allow surface to remain wet for the recommended contact time.

Because OPTIM 33 TB is approved as a One-Step Cleaner Disinfectant, depending on the soil load, the 2 step process may not be required due to this EPA claim.

Get Optim in our Store

VistaPure AutoFill

S3020
$208.05
In stock
1
Product Details
Brand: SciCan

VistaPure AutoFill

VistaPure’s patented AutoFill accessories replenish SciCan STATIM®and BRAVO™ autoclaves with distilled-quality water automatically — enabling truly hands-free convenience.

VistaPure AutoFill accessory kits work with all VistaPure systems and can be installed easily on any STATIM 2000/G4 or STATIM 5000/G4, or any BRAVO autoclave, in just a few minutes!

  • No more purchasing and storing distilled water
  • No more cycle-fault errors caused by forgetting to fill reservoirs
  • No more spills while filling autoclaves manually
  • No more staff time spent filling autoclaves = more time with patients

VistaPure AutoFill Installation

A 12-foot length of clear polyurethane tubing is already attached to the AutoFill manifold. If the desired distance between the connection to the VistaPure and the autoclave is greater than 12 feet, simply order a longer length of the clear polyurethane tubing, as the distance from the VistaPure to the autoclave is not limited, except by the feed tubing.

Installing AutoFill on a STATIM

The VistaPure AutoFill for STATIM device (S3020) can be installed on any STATIM 2000/G4 or STATIM 5000/G4 autoclave as follows:

https://www.scican.com/us/media/water-management/SD-491_US_EN%20R1_Autofill_Sell%20Sheet_20160805_PRNT.pdf

Installing AutoFill on a BRAVO

The VistaPure AutoFill for BRAVO device (S3030) can be installed on any model BRAVO autoclave.

In order to connect the VistaPure AutoFill to the BRAVO, one of the two tanks that come with the BRAVO must be selected as a “relay tank” since the BRAVO system uses a suction-lift pump to fill its reservoir when in its automatic-filling mode. Note that the relay tank must be located within 6 feet of the BRAVO unit, due to the suction-lift capacity of the BRAVO pump.


Connecting to a StatClave G4