OPTIM 33TB cleaning protocol

SciCan_OPTIM_USA

OPTIM™33 TB Quick Reference Guide

SWITCHING TO OPTIM 33 TB

Give all surfaces a rinse with mild soap & water before switching to OPTIM.

SciCan_OPTIM_USAQUATs (Quaternary Ammonium Compounds) or phenol based disinfectants are common active ingredients in the dental operatory today and are known to leave residual active chemistry on surfaces. AHP (Accelerated Hydrogen Peroxide) is an oxidizer and will remove chemical residues left by other surface disinfectants. There may be odor or film development if surfaces are not rinsed (damp wiped) properly with soap & water. This odor or film should however dissipate on average within 7 to 10 days of continual use as AHP will remove the residual active ingredient.

Surfaces need to be periodically rinsed with water!

Dissents or complaints happen when users are not informed on the correct protocol when using the product. It is good practice to periodically (at least once a week) rinse or wipe surfaces with a damp cloth to remove surfactants (detergents) or residual active ingredients that can build up over time with the use of ANY disinfectant cleaner.

No mixing of chemistries!

Mixing of chemistries in operatory i.e. using quats-alcohol/phenol disinfectant and OPTIM 33 TB in the same operatory or refilling OPTIM into chemical container used previously for other chemicals is not recommended. It is best practice to never mix chemistries (any kind) no matter what industry you are in.

On which surfaces can OPTIM 33 TB be used?

Non-porous hand contact environmental surfaces including Naugahyde (vinyl coated upholstery fabric).

NOTE: Soft surfaces such as chairs and cables should be periodically rinsed with water or wiped with a damp cloth to prevent the build-up of disinfectant or surfactants. As best practice, for any kind of disinfectant or cleaners used, chairs should be wiped with a damp cloth at the end of the day or at least once a week.

PRECAUTIONS:

– Surfaces that are composed of brass or copper, or other ferrous metals may show signs of discoloration or pitting with prolonged exposure.

– Anodized aluminum (often used on hand-pieces) and carbon tipped instruments should be avoided (i.e. burrs). Material impact can be reduced with rinsing or damp wiping.

We advise against wrapping gauze pad or ‘sleeves’ soaked in disinfectant on hand-pieces.


OPTIM 33 TB Solution

OPTIM 33TBSpray – Wipe – Spray Protocol

Visibly Soiled Surface – Clean before disinfecting

  1. Spray OPTIM 33 TB directly onto surface.
  2. Wipe visible soil with Stat-Wipes or clean paper towel to clean surface.
  3. Discard soiled towelette.
  4. Continue to disinfection step

Visibly Clean Surface – Disinfect pre-clean surface

  1. Spray OPTIM 33TB to cover the pre-clean surface. Allow the disinfectant to remain
  2. undisturbed on the surface for the contact time indicated on the label.
  3. *Do not overspray as this can create pooling of liquid. OPTIM 33TB is water based not alcohol based.
  4. Wipe with Stat-Wipes or clean paper towel.
  5. Discard used towelette.

*Using gauze dipped in solutions for wiping can cause pooling of liquid on surface. Wipe surface dry with paper towel if surface is still wet after desired contact time is achieved.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One- Step Cleaner Disinfectant. Therefore, OPTIM can be used on visibly clean surface. Depending on the soil load on the surface, the 2 step process is not required due to this EPA claim.


OPTIM 33 TB Wipes

Standard Protocol for Visibly Soiled Surface

Clean before disinfecting

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Clean surface with towelette to remove organic matter or visible blood.
  3. Discard used towellete.

Disinfect After Cleaning

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Disinfect surface with towelette.
  3. Allow surface to remain wet for the recommended contact time.
  4. Discard used towelette.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One-Step Cleaner Disinfectant. Depending on the soil load on the surface, users have the choice of cleaning the surface with one side of the wipe and disinfecting the surface with the reverse side of the wipes. Always allow surface to remain wet for the recommended contact time.

Because OPTIM 33 TB is approved as a One-Step Cleaner Disinfectant, depending on the soil load, the 2 step process may not be required due to this EPA claim.

Get Optim in our Store

Statim 2000 USED / REFURBISHED

2000SPECIAL
$3,760
Pre-order
1
Product Details
Brand: SciCan
Model: 2000 Classic

STATIM 2000 Cassette Autoclave


Reconditioned used units validated and spore tested with one year warranty

  • Ultra-fast 6 minute unwrapped cycle
  • 14 minute wrapped cycle
  • Exceptional instrument turnover
  • 10 minute Dri-Tecc Drying Technology
  • Aseptic transfer of instrument directly to the point-of-use
  • Reduced instrument investment
  • Prolongs instrument life

The STATIM 2000 Classic boasts a 6-minute cycle, but has the added advantage of employing a fully removable chamber so you can aseptically transport a set of freshly sterilized instruments directly to the procedure area.

The STATIM 2000 is fully automatic. Simply select one of the three sterilization programs and with a single touch of the keypad, begin the cycle. An advanced microprocessor controls all functions and maintains optimal sterilization conditions, from start to finish.

SciCan's STATIM cassette sterilizers have a biological effectiveness that has been proven in tests for hollow and solid instruments at many internationally recognized institutes in Canada, the United States, and Europe.

The STATIM 2000 is a reliable cassette sterilizer that's compact enough to fit in procedure areas and fast enough to sterilize jewelry, expensive Lasik, PRK, endoscopic and cataract instruments, including phaco handpieces, between procedures - it can complete an unwrapped sterilization cycle in just six minutes and a wrapped cycle in 14 minutes.

Dri-TecC Convection Heat Drying for the STATIM 2000

Dri-TecC – Convection heat drying technology

The STATIM 2000 uses convection heat to dry instruments by utilizing the remaining heat in the system after the sterilization phase. Heat is captured and released into the cassette, drying wrapped instrument loads in as little as 10 minutes.

Sterilization cycle description chart

Cycle Unwrapped Wrapped Rubber/Plastic
Sterilization time 3.5 minutes 10 minutes 15 minutes
Sterilization temp. 135º C 135º C 121º C
Total cycle time 6 minutes 14 minutes 20 minutes

ACCESSORIES available in the Shop:

01-100008A Cassette complete (with rack 01-106653)
01-100028S Seal & lubricant kit
01-100207S Air compressor filter
01-102119S Microbiological air filter
01-100812S Waste bottle complete
01-100204S Exhaust tube
01-103945S Instrument rack for unwrapped instruments
01-106653 Instrument rack for wrapped instruments
SCI134 Test Strip indicators (box of 250)
01-101783S Water reservoir cap & filter
01-109300S Water reservoir filter

statim_2000_cassettesysTM_cassetteSurgical Plastic Sterilization TraysStat-Dri Plus

SPECIFICATIONS:

Unit Size : L 480mm x W 415mm x H 150mm
L 19" x W 16.25" x H 6"
Cassette Internal Dimensions : L 280mm x W 180mm x H 35mm
L 11" x W 7" x H 1.5"
Reservoir Capacity : 4.0 litres (distilled water) Approximately 40 cycles
Weight without water : 21 kg/46 lbs
Power Consumption : 110-120 V, 60Hz, 1300 W

When you receive your Statim 2000 packing carton, the items listed below will be included.
If any of the items are missing, contact us immediately so that the situation can be corrected.

  • Cassette Tray and Lid
  • Unwrapped Instrument Rack
  • Waste Bottle
  • Bottle Lid Fitting
  • Power Cord
  • Operator's Manual
  • Exhaust Tube
  • Tube Mounting Hardware

Finance your equipment at our checkout

*The use of STATIM sterilizers was introduced to the body art field in the early 1990's by Brian Skellie at Piercing Experience, and developed in community along with other practical innovations in applied safety precautions and techniques.

Save this product for later

1 thought on “OPTIM 33TB cleaning protocol”

Leave a Comment

This site uses Akismet to reduce spam. Learn how your comment data is processed.