OPTIM 33TB cleaning protocol

SciCan_OPTIM_USA

OPTIM™33 TB Quick Reference Guide

SWITCHING TO OPTIM 33 TB

Give all surfaces a rinse with mild soap & water before switching to OPTIM.

SciCan_OPTIM_USAQUATs (Quaternary Ammonium Compounds) or phenol based disinfectants are common active ingredients in the dental operatory today and are known to leave residual active chemistry on surfaces. AHP (Accelerated Hydrogen Peroxide) is an oxidizer and will remove chemical residues left by other surface disinfectants. There may be odor or film development if surfaces are not rinsed (damp wiped) properly with soap & water. This odor or film should however dissipate on average within 7 to 10 days of continual use as AHP will remove the residual active ingredient.

Surfaces need to be periodically rinsed with water!

Dissents or complaints happen when users are not informed on the correct protocol when using the product. It is good practice to periodically (at least once a week) rinse or wipe surfaces with a damp cloth to remove surfactants (detergents) or residual active ingredients that can build up over time with the use of ANY disinfectant cleaner.

No mixing of chemistries!

Mixing of chemistries in operatory i.e. using quats-alcohol/phenol disinfectant and OPTIM 33 TB in the same operatory or refilling OPTIM into chemical container used previously for other chemicals is not recommended. It is best practice to never mix chemistries (any kind) no matter what industry you are in.

On which surfaces can OPTIM 33 TB be used?

Non-porous hand contact environmental surfaces including Naugahyde (vinyl coated upholstery fabric).

NOTE: Soft surfaces such as chairs and cables should be periodically rinsed with water or wiped with a damp cloth to prevent the build-up of disinfectant or surfactants. As best practice, for any kind of disinfectant or cleaners used, chairs should be wiped with a damp cloth at the end of the day or at least once a week.

PRECAUTIONS:

– Surfaces that are composed of brass or copper, or other ferrous metals may show signs of discoloration or pitting with prolonged exposure.

– Anodized aluminum (often used on hand-pieces) and carbon tipped instruments should be avoided (i.e. burrs). Material impact can be reduced with rinsing or damp wiping.

We advise against wrapping gauze pad or ‘sleeves’ soaked in disinfectant on hand-pieces.


OPTIM 33 TB Solution

OPTIM 33TBSpray – Wipe – Spray Protocol

Visibly Soiled Surface – Clean before disinfecting

  1. Spray OPTIM 33 TB directly onto surface.
  2. Wipe visible soil with Stat-Wipes or clean paper towel to clean surface.
  3. Discard soiled towelette.
  4. Continue to disinfection step

Visibly Clean Surface – Disinfect pre-clean surface

  1. Spray OPTIM 33TB to cover the pre-clean surface. Allow the disinfectant to remain
  2. undisturbed on the surface for the contact time indicated on the label.
  3. *Do not overspray as this can create pooling of liquid. OPTIM 33TB is water based not alcohol based.
  4. Wipe with Stat-Wipes or clean paper towel.
  5. Discard used towelette.

*Using gauze dipped in solutions for wiping can cause pooling of liquid on surface. Wipe surface dry with paper towel if surface is still wet after desired contact time is achieved.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One- Step Cleaner Disinfectant. Therefore, OPTIM can be used on visibly clean surface. Depending on the soil load on the surface, the 2 step process is not required due to this EPA claim.


OPTIM 33 TB Wipes

Standard Protocol for Visibly Soiled Surface

Clean before disinfecting

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Clean surface with towelette to remove organic matter or visible blood.
  3. Discard used towellete.

Disinfect After Cleaning

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Disinfect surface with towelette.
  3. Allow surface to remain wet for the recommended contact time.
  4. Discard used towelette.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One-Step Cleaner Disinfectant. Depending on the soil load on the surface, users have the choice of cleaning the surface with one side of the wipe and disinfecting the surface with the reverse side of the wipes. Always allow surface to remain wet for the recommended contact time.

Because OPTIM 33 TB is approved as a One-Step Cleaner Disinfectant, depending on the soil load, the 2 step process may not be required due to this EPA claim.

Get Optim in our Store

Statim 2000 G4

G4-121101
$6,490
Latest revision STATIM 2000 G4 for Piercers with new touchscreen, PC board, and probe bracket with recessed thermocouple
In stock
1
Product Details
Brand: SciCan
Model: 2000 G4

Fast, Gentle, Reliable, and Smart.

Introducing The New Generation of Statim, The Statim G4 Series

(Generation Four)

Download brochure

Canonsburg, PA – SciCan has set the benchmark for sterilization and instrument reprocessing technology.

For more than 25 years SciCan has been developing and manufacturing highly innovative infection control products that address the need for versatility, flexibility, and efficiency in today’s environments.

SciCan is proud to unveil the newest STATIM family member, the G4 series. The STATIM G4 series is the same renowned and trusted autoclave it has been for over 25 years but now boasts a new contemporary look, functionality and connectivity that is the first of its kind. The G4 technology will change the way you interact by providing a direct channel of communication through the Internet to increase productivity and limit downtime.

Still powered by SciCan’s signature steam technology to provide sterilization at speeds faster than conventional chambered autoclaves, the STATIM G4 series has been drastically upgraded with a level of interactivity not seen before.

New look

The STATIM G4 series has been modernized with a new fascia that is fresh and contemporary.

Color touch screen

A large 3.5” high-resolution touchscreen offers a vivid display of messages and current cycle information all with extraordinary clarity. The bubble level has been integrated into the high-resolution touchscreen.

Built-in data logger

Gone are the days of messy thermal paper, clumsy printers or potential failures with external data loggers. With increasing stringency amongst national guidelines, the STATIM G4 series now saves all the data from each cycle. This data can be copied onto the included USB drive, viewed on screen, or remotely stored in the Cloud via email.*

STATIM G4 technology

SciCan's STATIM G4 series Technology offers a platform with endless possibilities. The product expansion and modes of communication will provide visibility from every facet, from usability to troubleshooting.

This is a very exciting time for SciCan, as we are embarking into new territory with cutting edge technology. SciCan has always demonstrated innovation and this is just another example of how we are setting new standards for infection control.

*Based on your email providers storage capabilities


This model has all the benefits of our classic STATIM 2000 Cassette Autoclave

  • Ultra-fast 6 minute unwrapped cycle
  • 14 minute wrapped cycle
  • Exceptional instrument turnover
  • 10 minute Dri-Tecc Drying Technology
  • Aseptic transfer of instrument directly to the point-of-use
  • Reduced instrument investment
  • Prolongs instrument life

Like the STATIM 900, the STATIM 2000 boasts a 6-minute cycle, but has the added advantage of employing a fully removable chamber so you can aseptically transport a set of freshly sterilized instruments directly to the procedure area.

The STATIM 2000 is fully automatic. Simply select one of the three sterilization programs and with a single touch of the keypad, begin the cycle. An advanced microprocessor controls all functions and maintains optimal sterilization conditions, from start to finish.

SciCan's STATIM cassette sterilizers have a biological effectiveness that has been proven in tests for hollow and solid instruments at many internationally recognized institutes in Canada, the United States, and Europe.

The STATIM 2000 is a reliable cassette sterilizer that's compact enough to fit in procedure areas and fast enough to sterilize jewelry, expensive Lasik, PRK, endoscopic and cataract instruments, including phaco handpieces, between procedures - it can complete an unwrapped sterilization cycle in just six minutes and a wrapped cycle in 14 minutes.

Dri-TecC – Convection heat drying technology

The STATIM 2000 uses convection heat to dry instruments by utilizing the remaining heat in the system after the sterilization phase. Heat is captured and released into the cassette, drying wrapped instrument loads in as little as 10 minutes.

Sterilization cycle description chart

Cycle Unwrapped Wrapped Rubber/Plastic
Sterilization time 3.5 minutes 10 minutes 15 minutes
Sterilization temp. 135º C 135º C 121º C
Total cycle time 6 minutes 14 minutes 20 minutes


ACCESSORIES available in the Shop
: see below

01-100008A Cassette complete (includes 01-106653 Instrument rack for wrapped instruments)

01-100028S Seal & lubricant kit

01-100207S Air compressor filter

01-102119S Microbiological air filter

01-100812S Waste bottle complete (includes 01-100204S Exhaust tube)

SCI134 Test Strip indicators (box of 250)

SCI-T5 Sterilization Monitor Type 5 Chemical Indicator

01-101783S Water reservoir cap & filter (Includes 01-109300S Water reservoir filter)

SPECIFICATIONS:

Unit Size : L 480mm x W 415mm x H 150mm
L 19" x W 16.25" x H 6"
Cassette Internal Dimensions : L 280mm x W 180mm x H 35mm
L 11" x W 7" x H 1.5"
Reservoir Capacity : 4.0 litres (distilled water) Approximately 40 cycles
Weight without water : 21 kg/46 lbs
Power Consumption : 110-120 V, 60Hz, 1300 W

When you receive your Statim 2000 packing carton, the items listed below will be included.
If any of the items are missing, contact us immediately so that the situation can be corrected.

  • Cassette Tray and Lid
  • Instrument Rack
  • Waste Bottle
  • Bottle Lid Fitting
  • Exhaust Tube
  • Tube Mounting Hardware
  • Power Cord
  • Operator's Manual

STATIM G4 Cassette Autoclave In-service (United States)

An overview of installing, operating, and maintaining the SciCan STATIM G4 Cassette Autoclave


Save up to 35% off the list price of a STATIM 2000 or STATIM 5000 cassette autoclave when you use your Section 179 tax deduction: Click here for more info


*The use of STATIM sterilizers was introduced to the body art field in the early 1990's by Brian Skellie at Piercing Experience, and developed in community along with other practical innovations in applied safety precautions and techniques.

Save this product for later

1 thought on “OPTIM 33TB cleaning protocol”

Leave a Comment

This site uses Akismet to reduce spam. Learn how your comment data is processed.