SWITCHING TO OPTIM 33 TB

Give all surfaces a rinse with mild soap & water before switching to OPTIM.

SciCan_OPTIM_USAQUATs (Quaternary Ammonium Compounds) or phenol based disinfectants are common active ingredients in the dental operatory today and are known to leave residual active chemistry on surfaces. AHP (Accelerated Hydrogen Peroxide) is an oxidizer and will remove chemical residues left by other surface disinfectants. There may be odor or film development if surfaces are not rinsed (damp wiped) properly with soap & water. This odor or film should however dissipate on average within 7 to 10 days of continual use as AHP will remove the residual active ingredient.

Surfaces need to be periodically rinsed with water!

Dissents or complaints happen when users are not informed on the correct protocol when using the product. It is good practice to periodically (at least once a week) rinse or wipe surfaces with a damp cloth to remove surfactants (detergents) or residual active ingredients that can build up over time with the use of ANY disinfectant cleaner.

No mixing of chemistries!

Mixing of chemistries in operatory i.e. using quats-alcohol/phenol disinfectant and OPTIM 33 TB in the same operatory or refilling OPTIM into chemical container used previously for other chemicals is not recommended. It is best practice to never mix chemistries (any kind) no matter what industry you are in.

On which surfaces can OPTIM 33 TB be used?

Non-porous hand contact environmental surfaces including Naugahyde (vinyl coated upholstery fabric).

NOTE: Soft surfaces such as chairs and cables should be periodically rinsed with water or wiped with a damp cloth to prevent the build-up of disinfectant or surfactants. As best practice, for any kind of disinfectant or cleaners used, chairs should be wiped with a damp cloth at the end of the day or at least once a week.

PRECAUTIONS:

– Surfaces that are composed of brass or copper, or other ferrous metals may show signs of discoloration or pitting with prolonged exposure.

– Anodized aluminum (often used on hand-pieces) and carbon tipped instruments should be avoided (i.e. burrs). Material impact can be reduced with rinsing or damp wiping.

We advise against wrapping gauze pad or ‘sleeves’ soaked in disinfectant on hand-pieces.


OPTIM 33 TB Solution

OPTIM 33TBSpray – Wipe – Spray Protocol

Visibly Soiled Surface – Clean before disinfecting

  1. Spray OPTIM 33 TB directly onto surface.
  2. Wipe visible soil with Stat-Wipes or clean paper towel to clean surface.
  3. Discard soiled towelette.
  4. Continue to disinfection step

Visibly Clean Surface – Disinfect pre-clean surface

  1. Spray OPTIM 33TB to cover the pre-clean surface. Allow the disinfectant to remain
  2. undisturbed on the surface for the contact time indicated on the label.
  3. *Do not overspray as this can create pooling of liquid. OPTIM 33TB is water based not alcohol based.
  4. Wipe with Stat-Wipes or clean paper towel.
  5. Discard used towelette.

*Using gauze dipped in solutions for wiping can cause pooling of liquid on surface. Wipe surface dry with paper towel if surface is still wet after desired contact time is achieved.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One- Step Cleaner Disinfectant. Therefore, OPTIM can be used on visibly clean surface. Depending on the soil load on the surface, the 2 step process is not required due to this EPA claim.


OPTIM 33 TB Wipes

Standard Protocol for Visibly Soiled Surface

Clean before disinfecting

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Clean surface with towelette to remove organic matter or visible blood.
  3. Discard used towellete.

Disinfect After Cleaning

  1. Pull out OPTIM 33 TB towelette from the canister.
  2. Disinfect surface with towelette.
  3. Allow surface to remain wet for the recommended contact time.
  4. Discard used towelette.

** OPTIM 33 TB has been tested in accordance with EPA organic soil load requirement as a One-Step Cleaner Disinfectant. Depending on the soil load on the surface, users have the choice of cleaning the surface with one side of the wipe and disinfecting the surface with the reverse side of the wipes. Always allow surface to remain wet for the recommended contact time.

Because OPTIM 33 TB is approved as a One-Step Cleaner Disinfectant, depending on the soil load, the 2 step process may not be required due to this EPA claim.

Get Optim in our Store

Bravo G4 Chamber Autoclave

7A620010
$8,495
In stock
1
Product Details
Brand: SciCan
Model: Bravo 21V

The BRAVO G4 Chamber Autoclave is an EN13060 compliant Class B sterilizer that addresses a dental office’s need for speed, capacity, safety, and efficiency. Its cutting-edge sterilization technology incorporates both pre-sterilization vacuum to improve steam penetration and post-sterilization vacuum to improve drying.

Efficient

  • The BRAVO G4 chamber autoclaves offer efficient cycle times, sterilizing and drying instruments in as little as 33 minutes.
  • The delayed start feature allows the user to program cycles to begin at a later time; an effective feature when programming test cycles to begin before the start of the day.

Versatile

  • Offering enough capacity to meet the needs of any size practice, the BRAVO G4 comes in 17L, 22L, or 28L chamber size.
  • The BRAVO G4 offers numerous filling and draining options, including manual top and front filling, and automatic filling and draining through ports at the back.
  • The accessory water equipment and the BRAVO G4 footprint allows options for any Stericentre set-up.

G4 Technology

Connectivity & Data Logging

  • Integrated G4 technology in the BRAVO G4 allows for reliable load release, documentation and data storage.
  • WiFi enabled G4 technology can connect and send cycle data to your smart devices or computer, providing easily accessible cycle data and maintenance information.
  • To help optimise office processes, the BRAVO G4 can be configured to email cycle information or error messages directly to office staff and/or service technicians.
Cycle 17L 22L 28L
Cycle Type Cycle Time (mins) Max. Load (kg) Cycle Time (mins) Max. Load (kg) Cycle Time (mins) Max. Load (kg)
134°C Universal B 42 6 46 7.5 56 9
134°C Hollow Unwrapped S 35 6 39 7.5 44 9
134°C Solid Wrapped S 33 6 39 7.5 45 9
134°C Prion B 56 6 60 7.5 70 9
121°C Universal B 58 6 63 7.5 69 9
134°C Helix Test N/A 20 N/A 24 N/A 24 N/A
Vacuum Test N/A 18 N/A 18 N/A 18 N/A
Custom (134°C or 121°C) N/A N/A N/A N/A N/A N/A N/A


BRAVO G4 Sterilization, The Smarter Way.

The BRAVO G4, a Type B sterilizer, incorporates both pre-sterilization vacuum to improve steam penetration and post-sterilization vacuum to improve drying. The option of 17L, 22L or 28L units and efficient turnover times ensure clinicians always have an ample supply of instruments ready for use, regardless of practice size