STATIM Maintenance PDF Print E-mail
Autoclaves - STATIM

STATIM sterilizers should be cared for according to the manufacturer's recommendations

Download the Quick Reference Poster for your STATIM

GENERAL MAINTENANCE

To ensure trouble-free performance, operators must follow this preventative maintenance schedule.

Water Reservoir

Replace water as needed.

For ophthalmic use, drain at the end of every workday, leave empty, and refill at the start of the next workday.

Waste BottleWaste Bottle complete with tubing is included with the STATIM sterilizer

Empty the waste bottle every time you refill the water reservoir.

Fill the bottle with water, up to the MIN line marking. Get a new waste bottle or parts, such as tubing or the top fixture.

You may also add some chlorine-free disinfectant such as OPTIM.

CassetteSTATIM Cassettes

Wash the interior of the cassette with dishwashing soap or a mild detergent that does not contain chlorine.

Scrub the inside with a cleaning pad designed for use with Teflon®-coated surfaces.

STAT-DRI Plus dramatically shortens drying time in your autoclave!

Additional  STAT-DRI Plus can be ordered in our shop or by requesting 2OZPLUS, 8OZPLUST, or 32OZPLUS.

Water Reservoir Filter:

[Replacing the water reservoir filter: read full article]Water Reservoir Filter and Cap Kit

Check the water reservoir filter every week and clean if necessary. Replace Filter only if necessary.

IMPORTANT UPGRADE: Get a new cap and water filter kit if your STATIM does not have one.

Biological FilterAir and Biological Air Filters for STATIM 2000 (shown)

Replace every 500 cycles or six months (whichever is first). Get a new biological filter.

Call for service if wet.

 

Air FilterReplace your STATIM air filters every six months or else it will look like the gross dirty yellow one top left! (STATIM 2000 air compressor filter shown)

Replace every 500 cycles or six months (whichever is first). Get a new air filter.

 

Cassette Seal

[Seals go green! Read full article]Seal & Lube Kits

Replace every 500 cycles or six months (whichever is first), or whenever necessary. Get a new cassette seal.

 

The factory default setting is for the unit to work with a water filter installed. If you choose to use steam process distilled water instead of tap water, you can install the Water ByPass Cartridge (SCWFBP) and set up the unit to work accordingly.
To set up the unit to work with the Water ByPass Cartridge, follow these steps:
1. Remove the water reservoir and water filter as described previously and replace with the Water ByPass Cartridge.
2. Enter the User Menu (see section 3.4) and select the WATER FILTER option when it is displayed on the first line.
3. Using the (down) & (up) buttons, choose NOT INSTALLED, then press .
4. Exit the user menu by pressing .
To use the SciCan Water Filter Cartridge, you will need to enter the user menu and choose INSTALLED under the WATER FILTER option.

STATIM 7000 Water Filter Cartridge

 

Maintaining Fluid Levels

Unless you are using the SciCan Water Filter (SCWF1 Water Filter Cartridge), that allows the use of tap water, you should only use steam-process distilled water containing less than 5 ppm total dissolved solids (having conductivity of less than 10 ?S / cm) in the Statim. Each time you refill the reservoir, empty the waste bottle. Empty the waste bottle often to avoid unpleasant odors and discoloration of the contents. (A low-level chlorine-free disinfectant, prepared according to the manufacturer’s instructions, may be added to the waste bottle to remedy this situation). The condenser bottle should be cleaned and refilled to the MAX line on a weekly basis.

Changing The Water Filter

If you are using the SciCan Water Filter, it will need to be changed regularly to ensure clean water is entering the internal components of the unit. When the water filter cartridge is close to being exhausted, a message will be displayed when a cycle is selected and an audible tone reminds you to order or replace the filter. This message will appear for approximately two seconds and then disappear, and then the chosen cycle will start. This message will continue to appear every time a cycle is chosen until the water filter is replaced with a new filter. If the filter is not replaced and has been 100% used, the unit will not operate until a new filter is installed.The water filter has a shelf life and should be used prior to the expiry date inscribed on each filter.

When installed, the water filter has a maximum life of 60 days, depending on the quality of the tap water. The unit will notify you prior to the expiry or exhaustion of the filter.

The water filter can be found under the removable water reservoir.

To replace this filter, follow these steps:

1. Press the STOP button twice.

2. Remove the water reservoir.

3. Remove the used water filter and discard.

4. Remove the water filter from its packaging.

5. Insert the new water filter. The water filter should be placed flat on the bottom, with the tabs on the water filter facing upwards, and then pushed inward until the filter is fully seated along the bottom of the unit.

6. Replace the water reservoir.

At the beginning of the next cycle, the unit will prepare the new filter for about 2 minutes and display the message PLEASE WAIT.

 

The factory default setting is for the unit to work with a water filter installed. If you choose to use steam process distilled water instead of tap water, you can install the Water ByPass Cartridge (SCWFBP) and set up the unit to work accordingly.

To set up the unit to work with the Water ByPass Cartridge, follow these steps:

1. Remove the water reservoir and water filter as described previously and replace with the Water ByPass Cartridge.

2. Enter the User Menu (see section 3.4 of your manual) and select the WATER FILTER option when it is displayed on the first line.

3. Using the WRAPPED(down) & UNWRAPPED(up) buttons, choose NOT INSTALLED, then press RUBBER & PLASTICS.

4. Exit the user menu by pressing STOP.

To use the SciCan Water Filter Cartridge, you will need to enter the user menu and choose INSTALLED under the WATER FILTER option.

 

If you need further assistance, a STATIM user's manual or a Troubleshooting Guide, This e-mail address is being protected from spambots. You need JavaScript enabled to view it